85 • Bump Testing an Instrument with Special Sensors

If the bump test fails and the Force Calibration box was selected, the
SDM-E2 automatically performs a calibration immediately after the failed
bump test. The results screen will indicate the results of both the bump test
and calibration. The BUMP

T

LED on the SDM-E2 control panel will be

solid red. The CAL

S

LED will be solid green if the calibration passed or

solid red if the calibration failed. Click OK to return to the main program
window or double click an instrument information line to view the bump test
or calibration results in the Logs window. The BUMP

T

LED and CAL

S

LED on the SDM-E2 control panel will turn off.

Figure 63: Bump Test Failed/Calibration Failed