Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 192
10. Click OK. You will be asked if you want to bump test the instruments. Click OK
to proceed. If you don’t want to proceed with the bump test, click Cancel to
return to the main program window.
11. The pump in each instrument will start and the program will indicate that air is
flowing. The AIR indicator on the program screen will be on. The BUMP
T
LED on the SDM-E2 control panel will be flashing amber. Air will be applied to
the EAGLE 2 for 45 seconds, then the PC Controller Program will perform a
zero adjustment on all the channels.
Figure 135: Air Flowing
The PC Controller Program will analyze the zero adjustment results. If any
sensors failed the zero adjustment, the bump test will be aborted.