RKI Instruments SDM-E2 User Manual

Page 196

Advertising
background image

191 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-

If you are bump testing single channel special sensor instruments and have a
manifold assembled, the gas order confirmation screen will appear as shown
below. One sensor will be tested using GAS 1 (in this case, NH

3

). The

remaining sensors will be tested using GAS 2. The GAS 2 shown in the
column directly to the right of the GAS 1 column (in this case, IBL) will always
be bump tested before the GAS 2 shown in the following columns. The GAS 2
selection boxes are deselected by default. You must select the GAS 2 boxes if
you wish to bump test the special sensors. If you do not wish to bump test the
special sensors, leave the GAS 2 boxes deselected. Be sure that the
appropriate GAS 2 cylinder is connected to the GAS 2 fitting. The PC
Controller program will prompt you to change the GAS 2 cylinder during the
bump test.

Figure 134: Gas Order Confirmation Screen, Single Channel Units, Manifold

Advertising