RKI Instruments SDM-E2 User Manual
Page 82
Advertising
![background image](/manuals/571388/82/background.png)
77 • Bump Testing an Instrument with Special Sensors
11. The pump will start and the program will indicate that air is flowing. The AIR
indicator on the program screen will be on. The BUMP
T
LED on the SDM-E2
control panel will be flashing amber. Air will be applied to the EAGLE 2 for 45
seconds, then the PC Controller Program will perform a zero adjustment on all
the channels.
Figure 56: Air Flowing
The PC Controller Program will analyze the zero adjustment results. If any
sensors failed the zero adjustment, the bump test will be aborted.
Advertising