147 • Instrument Function Menu

A confirmation screen will appear asking if you would like to save the changes you
have made. To save the changes and return to the main program window, click Yes.
To return to the main program window without saving your changes, click No. To
return to the Parameter tab and update your changes, click Cancel.

To return to the main program window without saving any changes, click Cancel in
the lower right corner of the window.

Sensor Tab

The Sensor Tab displays the alarm settings and auto calibration values for all
installed sensors. It also shows the date and time of the most recent calibration for
each channel. The three lines at the bottom of the Sensor Tab that are displayed in
green are for special conditions. The first line, CH4(5000ppm), displays the alarm
points for the catalytic LEL channel while it’s detecting in the ppm range. The second
line, CH4(5.00vol%), displays the alarm points for %volume combustible gas
detection. The factory setting for these alarms is off (*****) but they can be set. The
third line, OXY(40.0vol%), shows the alarm points for the oxygen channel when the
EAGLE 2 is being used in Inert Mode. These oxygen alarms are both rising alarms
and will only appear if Inert Mode is factory set to On for the connected EAGLE 2.

Any value in a white box can be changed. Values in shaded boxes cannot be
changed. To change a value:

1.

Click the box whose value you want to change. The entire box will turn blue.

Figure 110: Sensor Tab