189 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-

9.

You will then be asked to confirm the gas cylinder concentration and the
instruments that the cylinder will be used for. The serial number and gas
combination of all selected instruments will appear sequentially beneath the
gas cylinder concentrations. The confirmation screen will indicate that the
calibration cylinder for the first special sensor (in this case, NH

3

) must be

connected to the GAS 2 fitting first and that you must change the GAS 2
cylinder midway through the bump test (in this example, to IBL and then to
CH

4

). The GAS 2 shown in the column directly to the right of the GAS 1

column will always be bump tested before the GAS 2 shown in the following
columns. Be sure that the appropriate GAS 2 cylinder is connected to the GAS
2 fitting. The screen below confirms that the standard 4 sensors will be bump
tested using GAS 1 and the special sensors will be bump tested using GAS 2.

The GAS 2 selection boxes are deselected by default. You must select the
GAS 2 boxes if you wish to bump test the special sensors. If you do not wish to
bump test the special sensors, leave the GAS 2 boxes deselected.

Figure 132: Gas Order Confirmation Screen