Bump Testing an Instrument with Special Sensors • 72

4.

Ensure that the gas out to EAGLE 2 line is connected to the inlet of the EAGLE
2.

5.

Select the instrument or instruments you wish to bump test and click on them
to highlight them as shown below. To select multiple instruments for bump
testing, hold down the Control button on the keyboard as you click them.

Figure 51: Instrument Selection

6.

If you would like an automatic calibration to occur after a failed bump test,
select the Force Calibration selection box.

7.

Click the Bump test button in the lower right corner of the main program
window.