RKI Instruments SDM-E2 User Manual

Page 91

Advertising
background image

Bump Testing an Instrument with Special Sensors • 86

If at any point during the bump test the flow rate to the instrument becomes
too low, the PC Controller will abort the bump test and display a failed
bump test screen. The BUMP

T

LED on the SDM-E2 control panel will be

solid red. Click OK to return to the main program window or double click the
instrument information line to view the bump test results in the Logs
window. The BUMP

T

LED on the SDM-E2 control panel will turn off.

Confirm all tubing connections are correct and that all lines are clear.

Figure 64: Low Flow Bump Test Failure

17. For all bump test result outcomes, the instruments will be displayed in the main

program window in Details View. To return the main program window to Icons
View, click the Icons radio button.

18. All bump test results can be viewed in the Logs window. See “Bump Test Data”

on page 134 for instructions for viewing bump test data.

Advertising