207 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations

If you are calibrating single channel special sensor instruments and do not
have a manifold assembled, the gas order confirmation screen will appear as
shown below. All sensors will be tested using GAS 1. Be sure that each
docking station has a demand flow regulator and the appropriate calibration
cylinder connected to the GAS 1 fitting.

Figure 148: Gas Order Confirmation Screen, Single Channel Units, No Manifold