Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 186

NOTE:

If you have an EAGLE 2 with an IR CO

2

sensor in the 0-10,000 ppm or

0-5.00 %vol range, be sure to replace the air filter at the AIR inlet fitting
with the CO

2

scrubber when testing an instrument with a CO

2

sensor

installed to ensure that the CO

2

present in fresh air is scrubbed out.

Remove the black end caps from each end of the scrubber before
installing onto the AIR inlet fitting. Replace the caps when the scrubber
is not in use. See the “Spare Parts List” on page 181 for a scrubber
with a tubing stub.

5.

Select the instruments you wish to bump test and click on them to highlight
them as shown below. To select multiple instruments for bump testing, hold
down the Control button on the keyboard as you click them.

Figure 129: Instrument Selection

6.

If you would like an automatic calibration to occur after a failed bump test,
select the Force Calibration selection box.

7.

Click the Bump test button in the lower right corner of the main program
window.