Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 204

7.

You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the
calibration. Leave the special sensors deselected in this screen.

Figure 145: Gas Cylinder Confirmation Screen

If you are calibrating multiple single channel special sensor instruments, the
selections you make in the gas cylinder confirmation screen are affected by
the layout of your system. If you do not have a manifold set up for your system,
connect each appropriate calibration cylinder to the GAS 1 fitting on the back
of the SDM-E2 that contains the matching special sensor and select all gases
to be applied through GAS 1. Each docking station will have its own demand
flow regulator and its own calibration cylinder. If you have a manifold set up for
your system, select only one gas to be applied through GAS 1. Any unselected
gases will be applied through GAS 2 and you will be prompted by the PC
Controller program to change the GAS 2 cylinder. Be sure the selected
calibration cylinder is connected to the GAS 1 manifold and that the calibration
cylinder for the first GAS 2 to be tested is connected to the GAS 2 manifold.