Logs Button • 118

Calibration Data

The EAGLE 2 is capable of saving calibration information for up to the 100 most
recent calibrations. When the maximum number of calibration files is reached, the
oldest calibration file is overwritten when a new one is saved. When an EAGLE 2 is
connected to the PC Controller Program and a complete download occurs, the
calibration files are retrieved and stored by the program. In addition, all calibrations
performed by the PC Controller Program are stored by the program. To view and
perform desired operations with the calibration files:

1.

While in the Logs window, find your instrument by serial number, then click the
expanded view symbol (+) of or double-click the serial number folder to view
the contents.

2.

Click the expanded view symbol (+) of or double-click the Calibration folder to
view the contents. Calibration data folders are arranged by year/month.

3.

Click the year/month folder you wish to view. All calibration data files for that
instrument in that particular year/month will appear in the upper right frame.
The file number, serial number, station ID, user ID, model, date/time, and
result of the calibration appear in the same line. If calibrations were performed
prior to connecting the EAGLE 2 with the SDM Docking Station PC Controller
software, data for those calibrations will not appear until a complete download
is performed.

Figure 87: Viewing Calibration Log Files