123 • Logs Button

operation. The serial number, user ID, station ID, and catalytic combustible channel
configuration that are entered in the instrument when it is turned on or that are
updated during operation are saved for the corresponding interval trend file. If the
instrument’s user ID or station ID are changed during operation, any interval trend
files that result from subsequent operating sessions will have the new user ID or
station ID saved in them. If the catalytic combustible channel gas configuration is
changed during operation using the relative response feature in Display Mode, the
new configuration will be saved in the data until it is changed or the unit is turned off.
Any catalytic combustible channel gas configuration changes made using the relative
response feature in Display Mode are cleared when the unit is turned off and the unit
returns to its original configuration when it is turned on.

The data may be viewed in both table format and graph format if at least 5 scheduled
data points have been logged at the programmed interval trend time. This does not
include events. If an interval data file has fewer than five scheduled data points, the
graph controls are not functional and the data is shown only in table format.

Interval trend data will not appear unless a complete download is performed. To view
and perform desired operations with the interval trend files:

1.

While in the Logs window, find your instrument by serial number, then click the
expanded view symbol (+) of or double-click the serial number folder to view
the contents.

2.

Click the expanded view symbol (+) of or double-click the Interval folder to
view the contents. Interval data folders are arranged by year/month.