Safety monitoring functions, Emergency stop, Chapter 9 – Rockwell Automation 440C Guardmaster 440C-CR30 Configurable Safety Relay User Manual User Manual

Page 45: Chapter

Advertising
background image

Rockwell Automation Publication 440C-UM001C-EN-P - November 2014

45

Chapter

9

Safety Monitoring Functions

Many types of safeguarding and safety devices and safety related signals can be
connected as inputs to the CR30. The Connected Components Workbench
(CCW) facilitates the selection and connection of the device. Each block is
assigned the next available settings for input terminals, test sources number of
inputs, pulse testing, discrepancy time, and input filter.

Emergency Stop

The Emergency Stop function block sets the parameters for typical emergency
stop push buttons. In the CCW, click and drag (or double-click) the block to an
available Safety Monitoring Function spot. When mechanical operated contacts
are used, these contacts must be direct-acting contacts.

Figure 30 - Emergency Stop Function Block

The available input selections for the Emergency Stop inputs are:

EI_00…EI_11 (embedded input terminals 00…11)

MP_12…MP_17 (multi-purpose terminals 12…17)

You can modify the number and types of inputs:

2 N.C.

2 OSSD

1 N.C.

Pulse testing can be disabled or set to 2 Sources. When 2 Sources is selected, the
next available test sources are automatically selected. You can modify the sources
afterward.

Advertising