Monitor profile, Spot check and triage profiles, Configure pulse rate alarms – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 102: Manual parameters frame

Advertising
background image

Note

You can specify pulse tone volume in the Pulse rate tab (located in the Alarms tab).

Monitor profile

Spot Check and Triage profiles

Configure pulse rate alarms

Follow these steps to set alarm limits for pulse rate.

1. Verify that you are using the Monitor profile, which contains the Alarms tab.
2. Touch the Alarms tab.
3. Touch the Pulse rate tab.
4. Enter the desired upper and lower alarm limits for pulse rate using the up/down arrow keys or

the keypad.

5. Touch the Home tab.

The new alarm settings display in the Alarm Limit control button.

Manual parameters frame

The Manual parameters frame, located in the lower right of the Home tab, supports manual entry of

parameters and displays measurements taken by some accessories.

Note

Manual parameters are not available in the Triage profile.

Note

Body mass index (BMI) is only available with an attached weight scale that

calculates BMI.

Note

When a measurement is transferred from an attached weight scale to the monitor,

the measurement displayed on the monitor is within one decimal place (0.1) of the

measurement displayed by the weight scale.

Note

You cannot manually enter temperature on a monitor configured with a SureTemp

Plus temperature module.

96 Patient monitoring

Welch Allyn Connex

®

Integrated Wall System