Monitor radio – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 118

Advertising
background image

Monitor radio

The monitor's radio operates on Welch Allyn FlexNet™ or other 802.11 networks.

Wireless network interface

IEEE 802.11 b/g, 802.11a

Frequency

802.11 b/g: 2.402 GHz to 2.4835 GHz

802.11a: 5.125 GHz to 5.875 GHz

Channels

Up to 14 in 802.11b/g, up to 24 in 802.11a; country-dependent

Security/encryption/

authentication

WPA2/AES (either EAP or PSK authentication)

Antenna

Internal multiband PIFA

Wireless data rates

802.11b: 1Mbps or higher during vitals transmission only

802.11a/g: 6Mbps or higher during vitals transmission only

(approximately 2 seconds per reading)

Agency approvals

US: FCC Part 15, Class B; C/UL; CE; 47 CFR Part 2.1093, 15.207, 15.209,

15.247, 15.407; FCC OET Bulletin 65C

Europe: CE; EN 50371; EN/ETSI 300 328 V1.7.1, 301 489-1 V1.6.1, 301

489-17 V1.2.1, 301 893 V1.4.1

Canada: RSS-210; RSS-GEN; RSS-102

Hong Kong: HKTA 1039

Protocols

UDP, DHCP, TCP/IP

Data transfer protocols

UDP/TCP/IP

Modulation

OFDM (802.11a/g), DSSS/CCK (802.11b)

Output power

40mW typical, country-dependent

Ancillary IEEE standards

802.11e, 802.11h, 802.11i, 802.11X

Channel restrictions in the 5-GHz band are determined by country.

Marking by the symbol ( ! ) indicates that usage restrictions apply. To ensure compliance with local

regulations, be sure the correct country in which the access point is installed is selected. This

product can be used with the following restriction(s):

France - Outdoor use is limited to 10 mW EIRP within the band 2454 to 2483.5 MHz.

Note

Effective Isotropic Radiated Power (EIRP).

Note

Some countries restrict the use of 5-GHz bands. The 802.11a radio in the monitor

uses only the channels indicated by the access point with which the radio

associates. The hospital IT department must configure access points to operate

with approved domains.