Power down the monitor, Reset the wall system – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 41

Advertising
background image

a wrench icon

and a system fault code to aid service personnel and engineers in

diagnosing the problem.

Power down the monitor

1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.

This power-down method, which places the monitor into Standby mode, ensures that patient

measurements are retained in the monitor memory for a maximum of 24 hours. (These saved

measurements are available for recall, printing, or to send electronically to the network.) This

method also ensures that any configuration settings you have changed and saved will be

maintained at the next startup.

Note

Because power is still available to charge the battery and power the monitor, the

monitor is in Standby mode.

Reset the wall system

If the wall system stops functioning, you can press and hold

for approximately 6 seconds

to allow the hardware to completely cycle off and to reset the wall system configuration

settings to the last saved default power-up configuration. The button is located on the front of

the wall system.

CAUTION Do not use a long press of

to power down the wall

system when it is functioning normally. You will lose patient data

and configuration settings.

Note

Because power is still available to charge the battery and power

the wall system, the wall system is in Standby mode.