Alarm messages
Priority
Alarm limit exceeded. NIBP diastolic HIGH.
High
Alarm limit exceeded. NIBP diastolic LOW.
High
Alarm limit exceeded. NIBP MAP HIGH.
High
Alarm limit exceeded. NIBP MAP LOW.
High
Alarm limit exceeded. Pulse rate HIGH.
Medium
Alarm limit exceeded. Pulse rate LOW.
Medium
Alarm limit exceeded. SpO2 HIGH.
High
Alarm limit exceeded. SpO2 LOW.
High
Alarm limit exceeded. SpHb HIGH.
High
Alarm limit exceeded. SpHb LOW.
High
Alarm limit exceeded. Temperature HIGH.
Low
Alarm limit exceeded. Temperature LOW.
Low
Technical alarms
Alarm messages
Priority
Low battery 5 minutes or less remaining.
High
Searching for pulse signal.
High
Communications module did not power on properly. Power down the device.
High
Network not found; check network cable connections.
Low
Powering down. Call for service.
Low
Battery is absent or faulty. Call for service.
Low
NIBP air leak; check cuff and tubing connections.
Low
NIBP not functional. Call for service.
Low
Unable to determine NIBP; check connections and tubing for kinks.
Low
Incorrect NIBP cuff size; check patient type.
Low
Inflation too quick; check NIBP cuff and tubing connections.
Low
Unable to determine NIBP; check inflation settings.
Low
Directions for use