Alarm messages

Priority

Alarm limit exceeded. NIBP diastolic HIGH.

High

Alarm limit exceeded. NIBP diastolic LOW.

High

Alarm limit exceeded. NIBP MAP HIGH.

High

Alarm limit exceeded. NIBP MAP LOW.

High

Alarm limit exceeded. Pulse rate HIGH.

Medium

Alarm limit exceeded. Pulse rate LOW.

Medium

Alarm limit exceeded. SpO2 HIGH.

High

Alarm limit exceeded. SpO2 LOW.

High

Alarm limit exceeded. SpHb HIGH.

High

Alarm limit exceeded. SpHb LOW.

High

Alarm limit exceeded. Temperature HIGH.

Low

Alarm limit exceeded. Temperature LOW.

Low

Technical alarms

Alarm messages

Priority

Low battery 5 minutes or less remaining.

High

Searching for pulse signal.

High

Communications module did not power on properly. Power down the device.

High

Network not found; check network cable connections.

Low

Powering down. Call for service.

Low

Battery is absent or faulty. Call for service.

Low

NIBP air leak; check cuff and tubing connections.

Low

NIBP not functional. Call for service.

Low

Unable to determine NIBP; check connections and tubing for kinks.

Low

Incorrect NIBP cuff size; check patient type.

Low

Inflation too quick; check NIBP cuff and tubing connections.

Low

Unable to determine NIBP; check inflation settings.

Low

Directions for use

Alarms 65