Guidance and manufacturer's declaration

EMC compliance

Special precautions concerning electromagnetic compatibility (EMC) must be taken for all medical

electrical equipment. This device complies with IEC EN 60601-1-2.

All medical electrical equipment must be installed and put into service in accordance with the

EMC information provided in this document.

Portable and mobile RF communications equipment can affect the behavior of medical

electrical equipment.

The Connex Integrated Wall System complies with all applicable and required standards for

electromagnetic interference.

It does not normally affect nearby equipment and devices.

It is not normally affected by nearby equipment and devices.

It is not safe to operate the monitor in the presence of high-frequency surgical equipment.

However, it is good practice to avoid using the monitor in extremely close proximity to other

equipment.

Emissions and immunity information

Electromagnetic emissions

The monitor is intended for use in the electromagnetic environment specified below. The customer or user of

the monitor should assure that it is used in such an environment.

Emissions test Compliance

Electromagnetic environment - guidance

RF emissions
CISPR 11

Group 1

The monitor uses RF energy only for its internal function. Therefore,

its RF emissions are very low and are not likely to cause any

interference in nearby electronic equipment.

RF emissions
CISPR 11

Class A

The monitor is suitable for use in all establishments other than

domestic establishments and those directly connected to the public

low-voltage power supply network that supplies buildings used for

domestic purposes, provided the following warning is heeded:

Harmonic

emissions
IEC 61000-3-2

Class A

119