Minimum alarm volume

Select the minimum alarm volume available. If you

select

High

, then

Medium

and

Low

are not

available to the clinician.
These controls are on the Alarms tab (on the

General tab).

Nurse call threshold

Select the minimum priority alarm that activates a

nurse call relay. If you select

High

, only high-level

alarms activate a nurse call relay.

Audio pause time

Specify the amount of pause time that is added to

the 60-second pause time. When a clinician pauses

an audio alarm tone, the tone is paused for the

combined amount of time.

SpO2 alarm condition delay

Specify the minimum amount of time that an SpO2

alarm condition must be active before audio and

visual signals occur.
SatSeconds is available with Nellcor SpO2 sensors.

If you select 0 seconds or 10 seconds, SatSeconds

is disabled, and it is removed from the SpO2 tab in

the Alarms tab.

SpHb alarm condition delay

Specify the minimum amount of time that an SpHb

alarm condition must be active before audio and

visual signals occur.

4. Do one of the following:

To continue in the Advanced Settings, touch another tab.

To exit the Advanced Settings and return to the Home tab, touch Exit.

Specify advanced display settings

1. Access the Advanced Settings.

a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.

2. Touch the Display tab.
3. Specify settings.

Setting

Action/Description

Display lock

Specify the required period of clinician inactivity

before the touchscreen locks.

Display power saver

Specify the required period of monitor inactivity

before the display turns off.
Clinician interactions, new vital sign

measurements, or alarm conditions automatically

turn on the display.

Device power down

Specify the required period of monitor inactivity

before the monitor turns off.

4. Do one of the following:

To continue in the Advanced Settings, touch another tab.

To exit the Advanced Settings and return to the Home tab, touch Exit.