Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 81

1. Properly size the blood pressure cuff and position it around the patient’s bare upper arm.
2. On the Home tab, touch
.
3. Select Automatic.
4. Use the numeric keypad to enter the length of time between NIBP measurements.
5. Touch Start intervals.
Note
Intervals are not available in all profiles. See the Profiles section for
more information.
Program intervals
You can configure the monitor to take automatic NIBP measurements at variable intervals. The
monitor comes with preset interval programs that can be edited to meet your needs. The numbers
below the program name indicate the length of time between each interval in the cycle.
Start program intervals
Follow these steps to configure the monitor to take automatic NIBP measurements at variable
intervals.
1. Properly size the blood pressure cuff and position it around the patient’s bare upper arm.
2. On the Home tab, touch
.
3. Select Program.
4. Touch the desired program.
5. Touch Start intervals.
Create a new program interval or edit an existing program
Follow these steps to create or edit a program interval.
1. On the Home tab, touch the interval button (
or
).
2. Select Program.
3. Touch the desired program.
Directions for use
Patient monitoring 75