Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 81

Advertising
background image

1. Properly size the blood pressure cuff and position it around the patient’s bare upper arm.

2. On the Home tab, touch

.

3. Select Automatic.
4. Use the numeric keypad to enter the length of time between NIBP measurements.
5. Touch Start intervals.

Note

Intervals are not available in all profiles. See the Profiles section for

more information.

Program intervals

You can configure the monitor to take automatic NIBP measurements at variable intervals. The

monitor comes with preset interval programs that can be edited to meet your needs. The numbers

below the program name indicate the length of time between each interval in the cycle.

Start program intervals

Follow these steps to configure the monitor to take automatic NIBP measurements at variable

intervals.

1. Properly size the blood pressure cuff and position it around the patient’s bare upper arm.

2. On the Home tab, touch

.

3. Select Program.
4. Touch the desired program.
5. Touch Start intervals.

Create a new program interval or edit an existing program

Follow these steps to create or edit a program interval.

1. On the Home tab, touch the interval button (

or

).

2. Select Program.
3. Touch the desired program.

Directions for use

Patient monitoring 75