Adjust vital sign alarm limits, Modify audio alarm notification – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 69

Advertising
background image

Adjust vital sign alarm limits

You can adjust vital sign alarm limits or turn off alarm limit checking for individual parameters.

WARNING Alarm limits are user adjustable. All alarm limit settings should take

into account the patient's condition and acute care needs. Appropriate alarm limits

should be set accordingly for each patient.

CAUTION Loss of power will cause the monitor to return to default settings. Each

time you power up the monitor, you must set alarm limits appropriate for your

patient.

1. On the Home tab, touch the alarm limits control in the selected parameter frame. For example,

to adjust the NIBP alarm limits, touch

.

2. Adjust vital sign alarm limits.

To adjust a limit: Enter the desired upper and lower alarm limits using the up/down arrow

keys or the keypad.

To turn alarm limits off or on for the vital sign: Touch

or

. This button

toggles to display the current alarm state.

If you turn off alarm limit checking for a vital sign, no visual or audio alarm signals will

occur for those limits. If alarm limit checking is off, the icon changes to

on the

Home tab in the parameter frame.

Modify audio alarm notification

You can modify the volume of all audio alarms.
Note

If the Allow user to turn off general audio option has been selected in Advanced

settings, you can turn off audio alarms, but turning off alarms is not recommended

in some circumstances, such as unattended monitoring.

WARNING The alarm volume should be loud enough for you to hear it from where

you are. Set the volume considering the environment and ambient noise levels.

As you are working in the Alarms tab, parameter measurements appear across the top of the tab.

1. Touch the Alarms tab.

Directions for use

Alarms 63