A title bar with a lock icon ( ) replaces the Navigation area at the bottom of the screen.

•

Patient information no longer appears at the bottom left of the screen.

•

All controls on the screen are locked, except for on the title bar. If you select any area
of the screen other than , a message appears.

Unlock the display

On the locked screen, touch and move Slide to unlock (located at the bottom right) to the

rightmost position on the slidebar.
The following occurs:
•

Patient information appears in the Patient frame.

•

The Navigation area appears.

•

Home tab controls are available for use.

The display also unlocks when any of the following occur:
•

An alarm condition.

•

An externally initiated action, such as taking or stopping an NIBP measurement or upgrading

software.

•

The monitor powers up.