Problems and solutions – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 150

Advertising
background image

Message

Possible cause

Suggested action

Functional error. Call for

service.

A critical configuration load error occurred Call for service.

No connection for send.

The monitor is not configured to the network Call for service.

Problems and solutions

The problems addressed in this table do not generate alarm or information messages on the

monitor.

Problem

Possible cause

Suggested action

No SpHb value is displayed

An SpO2-only cable is connected to the

monitor

Replace the SpO2-only cable

with an SpO2/SpHb (Masimo

Rainbow) cable.

The SpHb cable has expired

Note

A technical alarm appears.

Replace the SpHb cable.

Poor sensor placement on the patient

Remove the sensor from the

patient and reapply.

The monitor may have the SpHb license, but

the SpO2 module does not

Contact Welch Allyn to verify

that the SpO2 module

contains the SpHb license.

No weight measurement is

transferred from the scale to

the monitor

The scale is not connected

Inspect the USB cables from

the device to the adapter to

the scale to ensure that they

are connected properly.

The scale setting is incorrect

Ensure that the scale settings

are enabled for transfer.

144 Troubleshooting

Welch Allyn Connex

®

Integrated Wall System