Take a temperature in the Direct mode

Direct mode displays the temperature of the probe as long as the probe tip remains in place at the

measurement site and remains within the operating patient temperature range. The patient’s

temperature will reach final equilibrium in approximately 3 minutes at the oral and rectal

measurement sites and approximately 5 minutes at the axillary site.

The monitor enters Direct mode by the following methods.

After you complete a Predictive mode measurement, touch

to switch from Predictive

to Direct mode. The temperature frame (in the lower-left corner) changes to "MODE: Direct..."

as it switches to the Direct mode.

Remove the probe from the probe well, load a probe cover, select a temperature site, and

expose the probe to ambient air for more than 60 seconds to switch the monitor to Direct

mode. The temperature frame changes to "MODE: Direct...".

If you have a patient whose body temperature is below the normal temperature range and you

follow the previous step, the probe sensor identifies this condition and turns off the probe

preheater in order to accommodate the lower body temperature measurement.

WARNING Patient injury risk. Do not exceed the recommended temperature

measurement durations in Direct mode. Continuous measurement durations of 3

minutes at the oral and rectal sites and 5 minutes at the axillary site are

recommended for accurate measurement. Do not continuously measure beyond 10

minutes in any mode.

WARNING Inaccurate measurement risk. To ensure optimal accuracy, always

confirm that the correct mode and site are selected.

WARNING Patient injury risk. Prior to taking a temperature, instruct the patient

not to bite down on the probe as patient injury and damage to the probe may result.

CAUTION Probe covers are disposable, nonsterilized, and single-use. Probes are

also nonsterilized. Do not autoclave probes and probe covers. Ensure that probe

covers are disposed of according to facility requirements or local regulations.

1. Remove the temperature probe from the probe well.

The monitor sounds a tone as it enters the ready state.

2. Insert the probe into a new probe cover and press the probe handle down firmly.

3. Touch the Temperature site control

to choose from these measurement sites: oral,

pediatric axillary, or adult axillary.
The temperature frame changes to Direct mode approximately 60 seconds after the probe is

removed from the probe well.

The monitor sounds a tone to indicate the start of a Direct mode measurement.

4. Hold the probe tip in place at the oral or rectal measurement site for a total of 3 minutes and

for 5 minutes at the axillary site.

5. While the measurements are taking place, the temperature frame displays the patient's

continuous temperature measurements in degrees Fahrenheit and degrees Celsius.

Directions for use

Patient monitoring 83