Startup

Power

The power button, located on the front of the monitor, performs the following functions:

Powers up the monitor

Sets the monitor into Display power saving mode, except when an alarm condition is active

(brief press)

Resets the monitor and sets the monitor into Standby mode (press and hold for 6 seconds)

CAUTION Do not use a long press of

to power down the monitor when it is

functioning normally. You will lose patient data and configuration settings.

The LED in the center of the power plug symbol indicates the battery charging status:

Green indicates that AC power is present and that the battery is fully charged.

Amber indicates that AC power is present and that the battery is charging.

The monitor has distinct power states.

Monitor on

The monitor is operating on battery power or AC power. You can utilize the monitor's features, and

the display is active.

Display power saving

The monitor is operating on battery or AC power, but the display is off to conserve power. A brief

press of the power button sets the monitor into Display power saving mode from the active state.

Settings for this mode can be changed in the Advanced Settings Display tab.

Battery-powered accessories connected to the monitor continue to charge while the monitor is in

this mode and connected to AC power.

Note

The monitor will not enter the Display power saving mode while an alarm condition

is active. In addition, the monitor will exit this mode if an alarm occurs.