Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 115

Advertising
background image

NIBP specifications

Cuff pressure range

Meets or exceeds ANSI/AAMI SP10:2002 standards for cuff pressure

range

Systolic range

Adult: 30 to 260 mmHg
Pediatric: 30 to 260 mmHg
Neonate: 20 to 120 mmHg

Diastolic range

Adult: 20 to 220 mmHg
Pediatric: 20 to 220 mmHg
Neonate: 10 to 110 mmHg

Cuff Inflation Target

Adult: 160 mmHg (StepBP)
Pediatric: 120 mmHg (StepBP)
Neonate: 90 mmHg (StepBP)

Maximum Target Pressure

Adult: 280 mmHg (StepBP, SureBP)
Pediatric: 280 mmHg (StepBP, SureBP)
Neonate: 130 mmHg (StepBP)

Blood pressure determination time

Typical: 15 seconds
Maximum: 150 seconds

Blood pressure accuracy

Meets or exceeds ANSI.AAMI SP10:2002 standards for noninvasive

blood pressure accuracy (±5 mmHg mean error, 8 mmHg standard

deviation)

Mean Arterial Pressure (MAP) range
The formula used to calculate MAP

yields an approximate value.

Adult: 23 to 230 mmHg
Pediatric: 23 to 230 mmHg
Neonate: 13 to 110 mmHg

Pulse rate range (using blood pressure

determination)

Adult: 30 to 200 bpm
Pediatric: 30 to 200 bpm
Neonate: 35 to 220 bpm

Pulse rate accuracy (using blood

pressure determination)

±5.0% (±3 bpm)

Overpressure cutoff

Adult: 300 mmHg ±15 mmHg
Pediatric: 300 mmHg ±15 mmHg
Neonate: 150 mmHg maximum

SureTemp Plus temperature module specifications

Temperature range

80°F to 110°F (26.7°C to 43.3°C)

Calibration accuracy

±0.2°F (±0.1°C) (Direct mode)

Directions for use

Specifications 109