SpHb graphical trend view

The graphical trend view presents a trend of the real-time measurements over a user-selected

period. In Advanced settings, you can select the period displayed.

The graph shows total hemoglobin level on the y-axis and time on the x-axis (oldest measurements

on the left to newest measurements on the right). The entire graph updates every 10 seconds.

To the right of the graph, the frame displays the current measurement in numeric format.

Measure SpHb

WARNING Inaccurate measurement risk. Use only Masimo Rainbow SET sensors

and accessories on Masimo-equipped monitors.

WARNING The pulsations from intra-aortic balloon support can increase the

pulse rate displayed on the monitor. Verify the patient's pulse rate against the ECG

heart rate.

WARNING Patient injury risk. Do not attempt to reprocess, recondition, or recycle

any sensors or patient cables. Doing so might damage electrical components.

WARNING Pulse rate measurement might not detect certain arrhythmias

because it is based on the optical detection of a peripheral flow pulse. Do not use

the pulse oximeter as a replacement or substitute for ECG-based arrhythmia

analysis.

WARNING Use the pulse co-oximeter as an early warning device. As you

observe a trend toward patient hypoxemia, use laboratory instruments to analyze

blood samples to better understand the patient's condition.

Directions for use

Patient monitoring 93