Weight scale messages, Physical assessment instrument handles – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 146

Advertising
background image

Message

Possible cause

Suggested action

Retry temperature measurement.

Note

This message often

accompanies other

temperature messages.

A probe heater or data error

occurred

Retry the temperature

measurement. If the problem

persists, replace the probe.

User settings require

adjustment

Adjust the user settings and

retry.

Weight scale messages

Message

Possible cause

Suggested action

Weight scale not functional.

Call for service.

The weight scale is not operating properly. Call for service.

Physical assessment instrument handles

Symptom

Possible cause

Suggested action

The lamp does not illuminate

There is no lamp in the handle head Install a lamp in the handle head.

The lamp is burned out

Install a new lamp.

The other handle is off the cradle

Place the other handle in the

cradle.

The system is not powered up

Power up the system.

The platform handle controller

PCBA is defective

Call service.

The handle assembly is defective

Call service.

The lamp is too dim

The rheostat setting is too low

Increase the rheostat setting.

The platform handle controller

PCBA is defective

Call service.

The handle assembly is defective

Call service.

The lamp is too bright

The rheostat setting is too high

Decrease the rheostat setting.

The platform handle controller

PCBA is defective

Call service.

The handle assembly is defective

Call service.

The lamp brightness does not

adjust

The platform handle controller

PCBA is defective

Call service.

The handle assembly is defective

Call service.

The handle becomes very hot to the

touch

The lamp has been on for an

extended period of time

Return the handle to the cradle.

140 Troubleshooting

Welch Allyn Connex

®

Integrated Wall System