Patient data management messages, Radio messages – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 147

Advertising
background image

Patient data management messages

Message

Possible cause

Suggested action

Maximum number of patient

records saved. Oldest record

overwritten.

The maximum number of patient records in

the monitor's memory has been exceeded

On the Review tab, delete old

records to prevent the alarm

from appearing when new

records are saved.

Unable to access patient

information.

An error occurred when reading the patient

list or patient record during startup

Power down and restart the

monitor. If the error persists,

call for service.

No data to save.

No patient data is available

Take or enter vital signs

before saving.

Patient ID required to save

data.

The configuration requires a patient ID to

save data

Call for service.

Clinician ID required to save

data.

The configuration requires a clinician ID to

save data

Call for service.

Patient ID required to send

data.

The configuration requires a patient ID to

send data

Add a patient ID.

Patient list is full. Delete some

patients to add more.

The maximum number of patients was

exceeded

Delete a patient from the list

to add a new patient.

Stop intervals to select new

patient.

The monitor is set to take interval readings

Stop intervals before

changing the patient.

No connection for send.

No connectivity is available to support

sending data manually or automatically

sending data on manual save

Call for service.

Unable to retrieve list.

The monitor is unable to retrieve a patient

list from the network

Call for service.

Unable to identify clinician.

The clinician ID or password is incorrect

Confirm the clinician ID and

password (if applicable), and

retry.

Radio messages

Message

Possible cause

Suggested action

Radio not functional. Call for

service.

A hardware failure occurred (not currently

used)

Call for service.

The radio has the wrong software

The radio is not connected

Radio error. Power down and

restart.

The monitor and the radio failed to establish

communication with each other

Power down and restart. If

problem persists, call for

service.

Directions for use

Troubleshooting 141

Advertising