background image

Constant amber for low priority alarms

Home tab

Home tab notifications

Notification

Description

Device Status area

The area changes color and displays a message with an accompanying status

icon or button. If the alarm tone is in a pause interval, a timer countdown appears.
If multiple alarms and information messages are active, the Device Status area

shows the highest priority alarm. If the alarms are equal in priority, the most

recent alarm message appears. You can cycle through the messages for each

active alarm.

Parameter frame

The background color changes. Touch this area to pause or turn off an alarm

audio tone. Visual indicators and Nurse Call notification will persist during an

audio paused condition.

Alarm Limit control

The icon in this control indicates the status of the alarm limit settings. Red and

amber icons indicate measurements that have exceeded alarm limits.
Touch this control to navigate to a parameter-specific tab where you can modify

alarm limit settings.

Icons on the Home tab

Icons in parameter frames

The icons in the parameter frames indicate alarm notification settings. When alarm limits are on,

the icons will be black and white until an alarm occurs. Then, the icons will change color to

indicate the priority of the alarm. Red icons represent high priority alarms, and amber icons

represent medium or low priority alarms.