Temperature mode selection – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 86

Advertising
background image

WARNING Patient illness may result from improper use of oral and rectal

temperature probes. Using the incorrect probe may also produce inaccurate

measurements.

Use only oral probes, identified by a blue ejection button at the top of the probe, to

take oral and axillary temperatures.

Use only rectal probes, identified by a red ejection button at the top of the probe, to

take rectal temperatures.

WARNING Patient illness or cross-contamination may result from improper

placement of oral and rectal temperature probes in the probe wells.

Place only oral probes, identified by a blue ejection button at the top of the probe, in

the blue probe wells.

Place only rectal probes, identified by a red ejection button at the top of the probe, in

the red probe wells.

WARNING Inaccurate measurement risk. Never take an axillary temperature

through the patient's clothing. Carefully place the probe in the axilla, avoiding

contact with other objects or material. Always verify direct contact between the

probe cover and skin.

WARNING Patient injury risk. When taking rectal temperatures, insert the probe

tip only 5/8 inch (approximately 1.5 cm) inside the rectum of adults and only 3/8 inch

(approximately 1 cm) inside the rectum of children to avoid the risk of bowel

perforation.

WARNING Never use a damaged temperature probe. The thermometer consists

of high-quality precision parts and should be protected from severe impact or

shock. Do not use the thermometer if you notice any signs of damage to the probe

or monitor. If the thermometer probe is dropped or damaged, remove it from use

and have it inspected by qualified service personnel.

CAUTION Inaccurate measurement risk. Patient activities such as strenuous

exercise, ingesting hot or cold liquids, eating, chewing gum or mints, brushing teeth,

or smoking may affect oral temperature measurements for up to 20 minutes.

CAUTION Inaccurate measurement risk. Always use new probe covers taken

from the monitor's probe cover box holder to ensure accurate temperature

measurements. Probe covers taken from other places or that haven't stabilized in

temperature may result in inaccurate temperature measurements.

Temperature mode selection

The monitor with the temperature module takes a patient temperature in either Predictive (Normal)

or Direct mode. The default setting is the Predictive mode.

WARNING Inaccurate measurement risk. To ensure optimal accuracy, always

confirm that the correct mode and site are selected.

WARNING Patient injury risk. Do not exceed the recommended temperature

measurement durations in Direct mode. Continuous measurement durations of 3

minutes at the oral and rectal sites and 5 minutes at the axillary site are

recommended for accurate measurement. Do not continuously measure beyond 10

minutes in any mode.

80 Patient monitoring

Welch Allyn Connex

®

Integrated Wall System