WARNING Inaccurate measurement risk. Do not use the monitor on patients who

are experiencing convulsions or tremors.

WARNING Do not place the monitor in any position that might cause it to fall on

the patient.

WARNING Welch Allyn is not responsible for the integrity of a facility's power. If

the integrity of a facility's power or protective earth conductor is in doubt, always

operate the monitor on battery power alone when it is attached to a patient.

WARNING For operator and patient safety, peripheral equipment and

accessories that can come in direct patient contact must comply with all applicable

safety, EMC, and regulatory requirements.

WARNING All signal input and output (I/O) connectors are intended for

connection of only devices complying with IEC 60601-1, or other IEC standards (for

example, IEC 60950), as applicable to the monitor. Connecting additional devices to

the monitor may increase chassis or patient leakage currents. To maintain operator

and patient safety, consider the requirements of IEC 60601-1-1. Measure the

leakage currents to confirm that no electric shock hazard exists.

WARNING Equipment failure and patient harm risk. Do not cover the air intake

vents on the right or exhaust vents on the front of the Connex IWS. Covering these

vents could cause overheating or muffling of alarms.

WARNING This equipment is not suitable for use in the presence of electro-

surgery.

WARNING Cross-contamination or nosocomial infection risk. Clean and disinfect

the monitor on a routine basis according to your facility's protocols and standards

or local regulations. Thorough hand-washing before and after contact with patients

greatly reduces the risk of cross-contamination and nosocomial infection.

WARNING The physical assessment instruments (handles) are designed for

intermittent use. On-time should not exceed 2 minutes. Allow at least 10 minutes off-

time between patients.

CAUTION United States Federal law restricts this monitor to sale, distribution, or

use by or on the order of a physician or licensed healthcare professional.

CAUTION Welch Allyn is not responsible for the integrity of any wall mounting

interface. Welch Allyn recommends that you contact your Biomedical Engineering

Department or maintenance service to ensure professional installation, safety, and

reliability of any mounting accessory.

CAUTION Electromagnetic interference risk. The monitor complies with

applicable domestic and international standards for electromagnetic interference.

These standards are intended to minimize medical equipment electromagnetic

interference. Although this monitor is not expected to present problems to other

compliant equipment or be affected by other compliant devices, interference issues

still may occur. As a precaution, avoid using the monitor in close proximity to other

equipment. In the event that equipment interference is observed, relocate the

equipment as necessary or consult manufacturer's directions for use.

CAUTION Use only a Class I (grounded) AC power supply cord for powering this

monitor.