Braun ThermoScan PRO 4000 thermometer specifications (refer to manufacturer's
directions for use for additional information)

Temperature range

68°F to 108°F (20°C to 42.2°C)

Calibration accuracy

±0.4°F (±0.2°C) for temperatures ranging from 95.9°F to 107.6°F

(35.5°C to 42°C)

±0.5°F (±0.3°C) for temperatures outside of this range

Display resolution

0.1°F or °C

SpO2 specifications (refer to sensor manufacturer's directions for use for
additional information)

SpO2 performance measurement range

1 to 100%

Masimo sensor accuracy guide

Accuracy specified when used with Masimo SET pulse

oximetry monitors or with licensed Masimo SET pulse

oximetry modules using PC series patient cables,

during no motion. Numbers present ± 1 standard

deviation. Plus or minus one standard deviation

represents 68% of the population.

Perfusion

0.02 % to 20 %

Pulse rate

25 to 240 beats per minute (bpm)
No motion: ± 3 digits
Motion: ± 5 digits

Saturation

Note

Saturation accuracy varies by

sensor type.

70% to 100%
Adults, Pediatrics (No motion): ± 2 digits
Neonates (No motion): ± 3 digits
Adults, Pediatrics, Neonates (Motion): ± 3 digits
Low Perfusion: 0.02 % to 20 % ± 2 digits

Nellcor sensor accuracy guide

SpO2 measurement accuracy can only be evaluated in

vivo by comparing pulse oximeter readings with SaO2

measurements obtained from simultaneously sampled

arterial blood made using a laboratory CO-oximeter.

SpO2 accuracy was validated through breathe-down-

equivalent testing by Covidien using electronic

measurements to prove equivalence to the Nellcor

N600x predicate device. The Nellcor N600x predicate

device was validated by performing human-subject,

“breathe-down” clinical trials.

Perfusion

0.03 % to 20 %

Pulse rate

20 to 250 beats per minute (bpm) ± 3 digits

Saturation

Note

Saturation accuracy varies by

sensor type.

70% to 100%
Adult, Pediatrics: ± 2 digits
Neonate: ± 3 digits
Low Perfusion: 0.02 % to 20 % ± 2 digits

110 Specifications