CAUTION Do not use a long press of

to power down the monitor when it is

functioning normally. You will lose patient data and configuration settings.

CAUTION Never pull on the power cord when removing it from the power outlet.

When disconnecting the power cord, always grasp the attachment plug and not the

cord. Keep the cord away from liquids, heat, and sharp edges. Replace the power

cord if the strain relief or cord insulation is damaged or begins to separate from the

attachment plug.

CAUTION Use only the Welch Allyn USB client cable to connect a laptop

computer to the USB client port. Any laptop connected to the monitor must be

running on a battery, a 60601-1 compliant power supply, or a 60601-1 compliant

isolation transformer.

CAUTION If the touchscreen is not responding properly, refer to the

troubleshooting section. If the problem cannot be resolved, discontinue use of the

monitor and contact an authorized Welch Allyn service center or qualified service

personnel.

CAUTION Verify patient identity on the monitor after manual or barcode entry and

before printing or transferring patient records.

12 About warnings and cautions