Specifications, Physical specifications – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 113

Advertising
background image

Specifications

Physical specifications

Protection classifications, Wall system configurations

Characteristic

Specification

Electrical rating

100 – 240 V AC, 50 – 60 Hz, 1.5 – 0.8 A

Duty cycle - monitor

Continuous operation

Duty cycle - physical assessment

handles

2 minutes on, 10 minutes off

Type of protection against electric

shock

Class I equipment (protectively earthed) with double insulation

Degree of protection against electric

shock, for parts applied to patients

Type BF defibrillator proof
IEC EN 60601-1

Recovery time following defibrillator

discharge

Less than or equal to 10 seconds

Flammable anesthetics

WARNING Not suitable for use with flammable
anesthetics.

Degree of protection provided by the

enclosure with respect to harmful

ingress of liquids

IPX0
Non-protected according to EN/IEC 60529; Pulse oximeter equipment

complies with ISO 9919 Cl. 44.6 Ingress of liquids tests and EN/IEC

60601-1, 60601-2-30, 60601-2-49 Cl. 44.3 Spillage tests

Height

10.56 in. (268.26 mm )

Width

39.92 in. (1014 mm)

Depth

7.51 in. (190.8 mm)

Weight (including battery)

14.1 lb. (6 kg)

107