Specifications, Physical specifications – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 113

Specifications
Physical specifications
Protection classifications, Wall system configurations
Characteristic
Specification
Electrical rating
100 – 240 V AC, 50 – 60 Hz, 1.5 – 0.8 A
Duty cycle - monitor
Continuous operation
Duty cycle - physical assessment
handles
2 minutes on, 10 minutes off
Type of protection against electric
shock
Class I equipment (protectively earthed) with double insulation
Degree of protection against electric
shock, for parts applied to patients
Type BF defibrillator proof
IEC EN 60601-1
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Flammable anesthetics
WARNING Not suitable for use with flammable
anesthetics.
Degree of protection provided by the
enclosure with respect to harmful
ingress of liquids
IPX0
Non-protected according to EN/IEC 60529; Pulse oximeter equipment
complies with ISO 9919 Cl. 44.6 Ingress of liquids tests and EN/IEC
60601-1, 60601-2-30, 60601-2-49 Cl. 44.3 Spillage tests
Height
10.56 in. (268.26 mm )
Width
39.92 in. (1014 mm)
Depth
7.51 in. (190.8 mm)
Weight (including battery)
14.1 lb. (6 kg)
107