Message

Possible cause

Suggested action

Clear the alarm and retry

NIBP.

Incorrect NIBP cuff size; check

patient type.

The cuff size is not correct

Check the patient type.
Clear the alarm and retry

NIBP.

Inflation too quick; check NIBP

cuff and tubing connections.

NIBP inflation was too quick

Check the connections and

tubing for kinks.
Clear the alarm and retry

NIBP.

Unable to determine NIBP;

check inflation settings.

NIBP check inflation settings message

Check inflation settings and

change as necessary.
Clear the alarm and retry

NIBP.

Change the cuff inflation

target (CIT).

Excessive patient movement.

NIBP measurements are not accurate

because of artifact

Limit patient movement during

blood pressure measurement.

Tube type does not match

device configuration.
(NIBP measurement is

available)

The tube connected to the NIBP sensor does

not match the monitor's configuration

Use the tube specified for the

monitor.

Tube type does not match

device configuration.
(NIBP measurement is not

available)

User is using a single-lumen tube with the

following Advanced settings:
1.

Patient type is Pediatric or Adult

2.

Tube type is 2

3.

Algorithm is SureBP

Clear message. Modify

settings or tube use to match

patient type.

SpO2 and SpHb messages

Message

Possible cause

Suggested action

SpO2 not functional. Call for

service.

A module error has occurred

Try a new cable/sensor pair.
Call for service.

Searching for pulse signal.

(High-priority alarm)

The SpO2 sensor is not attached to the

patient's finger

Touch the alarm icon or the

SpO2 frame to dismiss the

alarm.

Set SpO2 alarm limits to OFF.

Reattach the SpO2 sensor to

the patient's finger.

Attach SpO2 sensor to monitor. The sensor was not detected

Check the sensor connection.

Replace the SpO2 sensor.

138 Troubleshooting