Cuff measurements – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 77

Advertising
background image

WARNING Never use an adult or pediatric monitor setting or cuff for an NIBP

measurement on a neonatal patient. Adult and pediatric inflation limits can be

excessive for neonatal patients, even if a neonatal cuff is used. Neonates are

defined in the AAMI SP10:2002 standard as children 28 days or less of age if born at

term (37 weeks gestation or more); otherwise, up to 44 gestational weeks.

CAUTION Correct sizing of the blood pressure cuff is important for accurate

blood pressure readings. A cuff that is too small might provide false high readings,

while a cuff that is too large might provide false low readings.

The monitor uses the oscillometric method to determine blood pressure; therefore, if the cuff

extends to the antecubital fossa (bend in the elbow), you can still acquire an accurate blood

pressure reading.

Before taking an NIBP measurement, follow these steps to select the appropriate cuff for the

patient.

1. Measure the circumference of the patient's bare upper arm, midway between the elbow and

shoulder.

2. Choose the appropriate cuff size based on the circumference measurement. If the

circumference of the patient's arm falls between two cuff sizes, use the larger cuff size.

3. Wrap the cuff around the patient's bare upper arm and verify that the artery index marker lies

somewhere between the two range markings on the cuff.

Cuff measurements

The following tables provide measurements for Welch Allyn blood pressure cuffs.

One-piece cuff measurements

Cuff Size

Circumference (cm)

Circumference (in)

Infant

9.0 – 13.0

3.5 – 5.1

Small child

12.0 – 16.0

4.7 – 6.3

Child

15.0 – 21.0

5.9 – 8.3

Small adult

20.0 – 26.0

7.9 – 10.2

Adult

25.0 – 34.0

9.8 – 13.4

Large adult

32.0 – 43.0

12.6 – 16.9

Thigh

40.0 – 55.0

15.7 – 21.7

Neonatal soft disposable cuffs with male Luer slips

Cuff Size

Circumference (cm)

Circumference (in)

NEO 1

3.3 – 5.6

1.3 – 2.2

NEO 2

4.2 – 7.1

1.6 – 2.8

Directions for use

Patient monitoring 71

Advertising