Troubleshooting, Nibp messages – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 143

Advertising
background image

Troubleshooting

This section presents tables of technical alarm and information messages, as well as problem

descriptions that do not generate messages, to help you troubleshoot issues on the monitor.
Note

Problem descriptions without messages appear at the end of this section.

When the monitor detects certain events, a message appears in the Device Status area at the top

of the screen. Message types include the following:

Information messages, which appear on a blue background.

Low- and medium-priority alarms, which appear on an amber background.

High-priority alarms, which appear on a red background.

Technical alarm messages are low priority unless noted in the Message column.

You can dismiss a message by touching the message on the screen, or, for some messages, you

can wait for the message to time out.

To use these tables, locate the message that displays on the monitor in the left column of the table.

The remainder of the row explains possible causes and suggests actions that can resolve the

issue.

Note

Instructions to "Call for service" in the following tables mean that you should contact

qualified service personnel in your facility to investigate the issue.

NIBP messages

Message

Possible cause

Suggested action

NIBP air leak; check cuff and

tubing connections.

The NIBP module has an air leak

Check the cuff and tubing

connections.
Clear the alarm and retry

NIBP.

NIBP not functional. Call for

service.

A module error occurred

Call for service.

Unable to determine NIBP;

check connections; limit patient

movement.

The NIBP module experienced a motion

artifact

Check connections; limit

patient movement.
Clear the alarm and retry

NIBP.

Unable to determine NIBP;

check connections and tubing.

The NIBP tubing has a kink

Check the connections and

tubing for kinks.

137