Maintenance and service

Perform periodic checks

Welch Allyn recommends that each facility conduct periodic checks of each monitor.

1. Check the following at least daily:

Audio (speaker and piezo beeper tones), especially at startup

Fan, especially at startup

Touchscreen alignment

Date

Time

2. Visually inspect the following at least weekly:

the monitor for any damage or contamination

all cables, cords, and connector ends for damage or contamination

all mechanical parts, including covers, for integrity

all safety-related labeling for legibility and adhesion to the monitor

all accessories (cuffs, tubing, probes, sensors) for wear or damage

documentation for current revision of the monitor

3. Visually inspect the following at least monthly:

Mounting screws on wall for looseness and wear

Update settings, replace items, or call for service as necessary based on results of visual

inspection. Do not use the monitor if you see any signs of damage. Qualified service personnel

must check any monitor that is damaged for proper operation before putting the monitor back into

use.

CAUTION Wall system components shall be replaced by Welch Allyn service

centers or qualified service personnel.

Remove the wall system from the wall

For maintenance or service activities that require access to the back of the wall system, follow

these steps to remove the wall system from the wall.

1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.