Battery power manager messages, Configuration manager messages – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 149

Advertising
background image

Message

Possible cause

Suggested action

persists, move the patient and

the monitor to a cooler

location.

Device shutdown is not

available at this time.

Device cannot perform an immediate

shutdown

Touch

OK

, wait, and retry.

Advanced settings unavailable Sensors are taking measurements

Stop continuous

measurements.

A physiological alarm condition is active

Respond to or reset the alarm.

Spot Check measurements have not been

saved

Save the measurements.

Unable to load language.

Chinese did not load

Power down and restart the

monitor.

Unexpected restart occurred.

Call for service.

A system error caused the monitor to restart. Call for service.

Battery power manager messages

Message

Possible cause

Suggested action

Low battery 5 minutes or less

remaining. (High-priority alarm)

Battery power is extremely low

Connect the monitor to AC

power. (If not connected to

AC power, the monitor powers

down when AC power is

depleted.)

Low battery 30 minutes or less

remaining.

Battery power is low

Touch the alarm icon to

dismiss or connect the

monitor to AC power.

Powering down. Call for

service.

Power manager or battery faults have

occurred

Call for service.

Battery is absent or faulty.

There is no battery in the monitor

Insert a battery.

The battery is faulty

Replace the battery.

Device is operating in battery

mode.

The AC power cord has been disconnected Touch

OK

to dismiss or

connect the monitor to AC

power.

Configuration Manager messages

Message

Possible cause

Suggested action

Unable to load configuration;

using factory defaults.

A configuration load error occurred

Call for service.

Directions for use

Troubleshooting 143