Position the wall system so that all instruments are accessible and in a location that allows for
ergonomic examinations.
Sample room layout
1
2
r
1.
Connex Integrated Wall System
2.
Examination table
Mount the wall system
1. On the selected wall, find and mark the studs, and choose the system height and
corresponding height for the mounting rail bracket.
Recommendation: Place the mounting rail bracket 63 in. (1.6 m) from the floor, which
places screen center height at approximately 63 in. (1.6 m) from the floor.
CAUTION This drawing shows the physical relationships of the
mounting brackets to each other and to the wall system after you
complete the mounting instructions. Do not place the wall system
on the wall until you have completed all preliminary steps.
2. Affix the mounting rail bracket to three studs at the selected height using the available screws
(anchors are provided for additional support).
CAUTION Ensure that the upper "lip" of the bracket sticks out
from the wall and that the bracket is level.
Directions for use
Setup 21