WARNING The accuracy of SpHb measurements can be affected by any of the

following:

elevated levels of total bilirubin

elevated levels of Methemoglobin (MetHb)

elevated levels of Carboxyhemoglobin (COHb)

hemoglobin synthesis disorders

low perfusion at the monitored site

the presence of concentrations of some intravascular dyes, sufficient to change the

patient's usual arterial pigmentation

patient movement

patient conditions such as shivering and smoke inhalation

motion artifact

painted nails

poor oxygen perfusion

hypotension or hypertension

severe vasoconstriction

shock or cardiac arrest

venous pulsations or sudden and significant changes in pulse rate

proximity to an MRI environment

moisture in the sensor

excessive ambient light, especially fluorescent

the use of the wrong sensor

a sensor applied too tightly

1. Verify that the sensor cable is connected to the monitor.

WARNING Patient injury risk. The sensor and extension cable are

intended only for connection to pulse co-oximetry equipment. Do

not attempt to connect these cables to a PC or any similar device.

Always follow the sensor manufacturer's directions for care and

use of the sensor.

2. Verify that you are using the Monitor profile.
3. Clean the application site. Remove anything, such as nail polish, that could interfere with

sensor operation.

Note

Do not use disposable sensors on patients who have allergic

reactions to the adhesive.

4. Attach the sensor to the patient according to the manufacturer's directions for use, observing

all warnings and cautions.

Note

If a sterile sensor is required, select a sensor that has been

validated for sterilization, and follow the sensor manufacturer's

directions for sterilizing the sensor.

Place the sensor and the NIBP cuff on different limbs to reduce unnecessary alarms when you

monitor these parameters at the same time.