4-3 Setting the Measurement Sensitivity

4-45

IG-E

4

Setting V

a

ri

o

us
F

u

nct
ion
s

When the filter value is increased, the measurement tends not to be influenced by dirt stuck
to the glass cover of the laser transmitter/receiver. When the filter value is decreased, small
objects are stably measured.
When the filter value in the "Standard mode" is 9, only if the light-blocking part of CCD
continues 9 pixels or more, it is judged as blocked. Also, only if the light-entering part of
CCD continues 9 pixels or more, it is judged not blocked.

Filter Value

Reference

• If the filter value is increased by 1, the light-blocking part and light-entering

part of CCD ignored will be increased by 1 pixel (approx. 0.014 mm).

• Depending on the light diffraction caused due to the mounting distance or

measurement position, the size of the object actually ignored changes.

(example) Filter value: 5

The blocked light only covers four pixels
and will be judged as received light.

The light blocked covers 5 pixels and will be
judged as a blocked area.

Receiver

Received light

1

2

3

4

CCD

Transmitter

Receiver

Received light

Received light

Blocked light

CCD