Addition mode, Addition mode -48 – KEYENCE IG Series User Manual

Page 106

Advertising
background image

4-4 Calculation Function

4-48

4

Setting V

a

ri

o

us
F

u

nct
ion
s

IG-E

In the addition mode, the added value of the internal measurement value (R.V.) for the
"main unit (A)" and the "expansion unit 1 (B)" is considered as the calculation value (CALC
value).

Calculation value = (A’s R.V.) + (B’s R.V.)

The diameter or width of a large object that cannot be measured with the measurement
range of one sensor head can be measured.

Setting example 1

If the zero shift function is used, the actual
diameter or width of the object can be displayed
as a calculation value (CALC value).

"3-7 Zero Shift Function (Shifting the Internal

Measurement Value (R.V.))" (page 3-16)

Setting example 2

If the zero shift function is used, the actual
diameter or width of the object can be displayed
as a calculation value (CALC value).

"3-7 Zero Shift Function (Shifting the Internal

Measurement Value (R.V.))" (page 3-16)

Addition Mode

Item

Main unit

Expansion

unit

1. Measurement

mode

Edge control/Positioning

2. Measurement

direction

Bottom

Top

Item

Main unit

Expansion

unit

1. Measurement

mode

Edge control/Positioning

2. Measurement

direction

Top

Main unit

Expansion unit 1

Mounting
bracket
(Optional)

Advertising