Pqw, 5*k(6, E/.qe – KEYENCE IG Series User Manual

Page 158: Append ix ig-e, Display description corrective actions

Advertising
background image

A-8

Append

ix

IG-E

The judgment
value (P.V.) is dis-
played as "

".

The hold function is used and
the judgment value (P.V.) is
held.

Check the hold function setting.
Input the timing input according to the setting.

"5. Hold function" (page 4-20)

The reset input is ON.

Set the reset input to OFF.
Check the wiring of the reset input.

"11. External input" (page 4-33)

The internal mea-
surement value (R.V.)
is displayed as "



HHHH".

The internal measurement
value (R.V.) falls below the
lower limit (-99.999) of display
range.

• Check the setting for shift target value and shift

the display again.

"3-7 Zero Shift Function (Shifting the
Internal Measurement Value (R.V.))"
(page 3-16)

• Redo the calibration.

"4-5 Calibration Function" (page 4-52)

The internal mea-
surement value
(R.V.) is displayed
as "

HHHH".

The internal measurement
value (R.V.) goes beyond the
upper limit (99.999) of display
range.

The key is locked and the but-
ton operation is disabled.

Cancel the key lock (Unlock).

"3-9 Key Lock Function" (page 3-21)

The tolerance tuning was per-
formed when the R.V. is
[

].

Perform the tuning when the internal measure-
ment value (R.V.) is displayed with figures (with
the object measured).

"3-6 Setting the Tolerance Setting Value"
(page 3-11)

When the R.V. is [

], the

zero shift function was used.

Shift the display when the internal measurement
value (R.V.) is displayed (with the object mea-
sured).

"3-7 Zero Shift Function (Shifting the Inter-
nal Measurement Value (R.V.))" (page 3-16)

The SET1 or SET2 value is
incorrect and the correction
using the calibration function
could not be performed.

Perform the calibration within the possible cor-
rection range.

"4-5 Calibration Function" (page 4-52)

The button operation is dis-
abled.

The connected communication unit <DL-RS1A>
read/write setting switch is set to RW.
Set the read/write setting switch to R.

"RS-232C Communication Unit DL-RS1A
User's Manual"

Display

Description

Corrective actions



LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



HOLD

CALC

CHECK

GO

((((

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

((((

HOLD

CALC

CHECK

GO

((((

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

((((

HOLD

CALC

CHECK

GO

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

.QE

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

PQW#.

GO

HOLD

CALC

CHECK

'TT

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

5*K(6

GO

HOLD

CALC

CHECK

'TT

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

5'6

GO

HOLD

CALC

CHECK

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

E/.QE

GO

HOLD

CALC

CHECK

Advertising