3-6 Setting the Tolerance Setting Value
3-14
3
B
as
ic Oper
a
tion
s
IG-E
Two-point tuning
With this method, the median value of the good target and defective target is set as the
tolerance setting value when there are good target and HIGH/LOW defective target.
1
Press the
W / X button several times on the main screen. Then display the
HIGH side setting value on the sub display (lower level).
"Sub Display (Lower Level)" (page 3-4)
2
Measure the good target and press the [SET] button. (HIGH side 1st point
confirmation operation)
The internal measurement value (R.V.) is imported as a good target measurement
value.
[
JKUGV] is displayed on the main display (upper level).
3
Measure the HIGH side defective target and press the [SET] button. (HIGH
side 2nd point confirmation operation)
The internal measurement value (R.V.) is imported as a measurement value for
HIGH side defective target.
After [
UGV] blinks on the main display (upper level) , the judgment value (P.V.) is
displayed.
On the sub display (lower level), the median value of the good target value (1st point)
imported on step 5 and HIGH defective target value (2nd point) is displayed.
Setting the HIGH side setting value (upper limit) is complete.
4
Press the
X button once and display the LOW side setting value on the sub
display (lower level).
Reference
• When the internal measurement value (R.V.) is [
], it is regarded as
0.000.
• When the internal measurement value (R.V.) is [
HHHH] and [HHHH], they are
regarded as -100.00 and 100.00 respectively.
• When the tuning result exceeds the setting range (-99.999 to 99.999), the
limit value of the setting range becomes the setting value.
GO
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
HOLD
CALC
CHECK
[HI]
ON
HIGH side setting value
GO
HOLD
CALC
CHECK
LASER
BANK
0
1
2
3
HI
LO
R.V.
ANALOG
HI
SHIFT
ZERO SHIFT
TIMING
LO
ALIGNMENT
LOW side setting value
[LO]
ON