Setting the shift target value, Setting the shift target value -16 – KEYENCE IG Series User Manual

Page 52

Advertising
background image

3-16

IG-E

3

B

as
ic Oper

a

tion

s

3-7

Zero Shift Function (Shifting the Internal Measurement Value (R.V.))

The internal measurement value (R.V.) is shifted (offset) to an arbitrary shift target value.
The judgment value (P.V.) is shifted (offset) as well.
The following two methods can be used.
• Press the [ZERO SHIFT] button (within 1 second).
• Set the external input (zero shift input) to ON for 20 ms or more.*

* When the zero shift input is set for the external input 1 (pink wire), the zero shift is

enabled by the external input.

"11. External input" (page 4-33)

1

Press the

W / X button several times on the main screen. To display the

shift target value on the sub display (lower level).

"Sub Display (Lower Level)" (page 3-4)

2

Press

S / T button to set the shift target value.

3

Press

W / X button to return the sub display to the original screen.

Setting the Shift Target Value

Item

Setting range

Default value

Shift target value

-99.999 to 99.999

0.000

GO

HOLD

CALC

CHECK



LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT



Shift target value

[SHIFT]

ON