Measurement direction, Average number of times (response time) – KEYENCE IG Series User Manual

Page 75

Advertising
background image

4-2 Basic Settings and Advanced Settings

4-17

IG-E

4

Setting V

a

ri

o

us
F

u

nct
ion
s

This is displayed when any of "Edge control/Positioning mode", "Glass edge mode" or "Pin
position measurement mode" is set for "1. Measurement mode" (page 4-8). (If
measurement modes other than the above are set, this screen is skipped.)

For details, refer to the "1. Measurement mode" (page 4-8).

The response time is the time from when the sensor head starts the measuring operation
and the value at the measurement position is established as an internal measurement
value (R.V.) and judgment value (P.V.), to time until when the judgment is output. When the
average number of times is increased, the response time becomes longer, and the internal
measurement value (R.V.) and judgment value (P.V.) stabilize.

*1 When "Pin interval judgment mode" or "Pin diameter judgment mode" is set for "1.

Measurement mode" (page 4-8),

"JUR" and "" to "" cannot be set.

Response time list

*1 Response time when modes other than "Pin interval judgment mode" and "Pin

diameter judgment mode" are set for " 1. Measurement mode" (page 4-8)

2. Measurement direction

Item

Setting range

Default value

Measurement Direction

VQR (Top), DVO (Bottom)

VQR

3. Average number of times (Response time)

Item

Setting range

*1

Default value

Average number of times

JUR /  /  /  /  /  /  /  /  /  / 
/

 /  / 



Configuration

Average

number of

times

Sampling cycle

(ms)

R.V. update cycle

(ms)

Response time

(ms)

*1

hsp

1

0.49

0.49

1.96

1

1

0.98

0.98

3.92

2

2

4.90

4

4

6.86

8

8

10.78

16

16

18.62

32

32

34.30

64

64

65.66

128

128

128.38

256

256

253.82

512

512

1.96

505.68

1024

1024

3.92

1009.40

2048

2048

7.84

2016.84

4096

4096

15.68

4031.72

Reference

The same values as the above are assigned for the analog output.