Specified edges interval measurement mode, 2 basic settings and advanced settings, Specifying the edge – KEYENCE IG Series User Manual

Page 74: Edge number setting screen, Reference, Setting v a ri o us f u nct ion s ig-e, Average number of times, 2 display screen and output" (page a-9), Ab c, Mode] or x [mode] or x

Advertising
background image

4-2 Basic Settings and Advanced Settings

4-16

4

Setting V

a

ri

o

us
F

u

nct
ion
s

IG-E

Specified edges interval measurement mode

The internal measurement value (R.V.) is determined by
measuring the distance from one edge to the other edge
of the target. Specify the two edges for measurement.

„ Specifying the edge

To specify the edge, when counting from the top, use a positive number (1 to 99). When counting
from the bottom, use a negative number (-1 to -99).
When specifying the edge of the measurement range on the top side, specify the top. When
measuring from the edge of the measurement range on the bottom side, specify the bottom.
(example) Setting for measuring the length of A, B and C with 2 pins of the object.

* The bracket ( ) shows the edge number when counting edges from the bottom side.

„ Edge number setting screen

When the specified edges interval measurement mode is set, the display goes to the edges number setting screen.

Press

S / T button on the edge number setting screen to specify the edge number.

Specify the edge for the top side on the edge number setting screen 1 and set the edge for
the bottom side on the edge number setting screen 2. If the setting is made reversely, a
negative value will be assigned to the internal measurement value (R.V.).

Specified range

Two edge numbers to specify

A

Either Top and 1, or Top and -4

B

Any one of 1 and 2, 1 and -3, -4 and 2, or -4 and -3

C

Any one of 1 and 4, 1 and -1, -4 and 4, or -4 and -1

Reference

If there is no specified edge within the measurement range, the internal
measurement value (R.V.) is determined as [

].

"2 Display Screen and Output" (page A-9)

Edge number

A

B

C

Top

Bottom

Top

1 (–4)
2 (–3)
3 (–2)
4 (–1)

Bottom

Item

Setting range

Default value

Edge number

99 to 1,

VQR (Top), DVO (Bottom), -1 to +99

1 or 2

* 75T

LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

 (PE

GO

HOLD

CALC

CHECK



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

'F)

HOLD

CALC

CHECK



LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

'F)

HOLD

CALC

CHECK



LASER

BANK

0

1

2

3

HI

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

 #W'

GO

HOLD

CALC

CHECK

1. Measurement mode

Edge number setting screen 1

Edge number setting screen 2

[MODE]
or

X

[MODE]
or

X

3. Average number of times

[MODE]
or

X

Advertising